5-2
Calibration
Mounting Details
Single Transducer Mounting Detail for Calibration
Single transducer calibration requires the use of a reaction
stand and transducer calibration block.
As shown in Figure 5-2, the transducer is secured to the
calibration block with two quick release pins.
Both pins must be installed for safety and stability of the
mount.
The white mark on the transducer is aligned in the up
position.
Figure 5-2: Single Low Torque Transducer Mounting Detail
Torque Under 250 ft lb
Mount the calibration block at the center of the stand with two
1/2-13 socket head cap screws with test bars that are 10" or
shorter. Install these screws using the supplied torque wrench
and 3/8-3/8 socket driver adapter. Use this configuration for up
to 250 ft lb of applied torque.
A 1/2" to 3/8" reducer is required when mounting the
10" arm to transducers having 3/8" drives.